NETEC Resource Library

Bioaerosol sampling for the detection of aerosolized influenza virus

Item

Click for External Resource*


Click to read full article*


*The link above may share a zip file (.zip) hosted on repository.netecweb.org. Zip files will download automatically.
*All other links are external and will open in a new window. If you click an external link, you are leaving the NETEC site, and we do not maintain, review, or endorse these materials. See our terms of use.


Files for Download

Blachere_et_al-2007-Influenza_and_Other_Respiratory_Viruses.pdfBlachere_et_al-2007-Influenza_and_Other_Respiratory_Viruses.pdf

Item Type

Publication

Terms of Use

By accessing these materials you are agreeing to our terms of use, which may be found here: Terms of Use.

Document Viewer

Was this resource helpful?


Title

Bioaerosol sampling for the detection of aerosolized influenza virus

Subject

Description

A Collison single-jet nebulizer was used to aerosolize the attenuated FluMist vaccine into a calm-air settling chamber.

Date

2007-05

Citation

Blachere, Francoise M., William G. Lindsley, James E. Slaven, Brett J. Green, Stacey E. Anderson, Bean T. Chen, and Don H. Beezhold. 2007. "Bioaerosol sampling for the detection of aerosolized influenza virus." Influenza and Other Respiratory Viruses 1 (3):113-20.

Abstract

Abstract

BACKGROUND:

Influenza virus was used to characterize the efficacy of a cyclone-based, two-stage personal bioaerosol sampler for the collection and size fractionation of aerosolized viral particles.

METHODS:

A Collison single-jet nebulizer was used to aerosolize the attenuated FluMist vaccine into a calm-air settling chamber. Viral particles were captured with bioaerosol samplers that utilize 2 microcentrifuge tubes to collect airborne particulates. The first tube (T1) collects particles greater than 1.8 microm in diameter, while the second tube (T2) collects particles between 1.0 and 1.8 microm, and the back-up filter (F) collects submicron particles. Following aerosolization, quantitative PCR was used to detect and quantify H1N1 and H3N2 influenza strains.

RESULTS:

Based on qPCR results, we demonstrate that aerosolized viral particles were efficiently collected and separated according to aerodynamic size using the two-stage bioaerosol sampler. Most viral particles were collected in T2 (1-1.8 microm) and on the back-up filter (< 1 microm) of the bioaerosol sampler. Furthermore, we found that the detection of viral particles with the two-stage sampler was directly proportional to the collection time. Consequently, viral particle counts were significantly greater at 40 minutes in comparison to 5, 10 and 20 minute aerosol collection points.

CONCLUSIONS:

Due to a lack of empirical data, aerosol transmission of influenza is often questioned. Using FluMist, we demonstrated that a newly developed bioaerosol sampler is able to recover and size fractionate aerosolized viral particles. This sampler should be an important tool for studying viral transmission in clinical settings and may significantly contribute towards understanding the modes of influenza virus transmission.

Accessibility

Free online (open access)

Collection